Troubleshooting Guide
Problem
1
The Display measurement shows
**** after a sample
2
The PWR lights do not light up
after switching on
The Display shows BATT
3
The Display remains black after
4
switching it on
6
The device is connected to the
power supply, but the BT light
does not light up
The device is connected to the
7
Computer, but I cannot find it
8
The measured value is not saved
9
The log file shows a wrong date
The log file shows a wrong time
10 The measured value varies
between two measurements
11 The device is not measuring
Solution
Value is higher than 5% oxygen. You cannot measure
oxygen in fresh air (21%).
Does the display show the main screen?
The LED light may be broken, but unless there is an
error on the LCD in most cases the device is not
charged.
Battery is uncharged.
Charge the device and try to switch it on again
Did you disconnect your device properly from the
Computer the last time?
Yes: The battery might be empty.
No/ I do not know: Contact CO2Meter.
If the device is switched off while connected to a power
outlet, the BT light does not light up while the device is
charging.
You must switch on the device before connecting to the
computer. If the device is switched on but you cannot
find it, check for available COM ports.
Did you press the LOG button? Did you observe the LOG
sign on the display? If yes, contact CO2Meter.
Check the date and time on the device is set. To
change date/time see the manual.
Give the TecPen time to acclimatize between different
rooms or temperatures.
Example: Can you hear the pump?
Yes: Did you follow operation procedure?
Yes: Contact CO2Meter.
Is the display showing MEAS?
No: Start the measurement.
Yes: Check screen 4 if there is a value of amplitude
higher than 0. If yes, contact CO2Meter.com
12