Coaguchek XS System Getting Started

Browse online or download pdf Getting Started for Blood Glucose Meter Coaguchek XS System. Coaguchek XS System 3 pages.
Also for Coaguchek XS System: Getting Started (2 pages), How To Use Manual (2 pages), Manual (17 pages)

Coaguchek XS System Getting Started
CoaguChek
System
This is a CLIA waived system.
SETTING DATE AND TIME
1. Start Setup
To set the date and time, you'll use
these buttons:
to change a setting
or
to accept a setting.
If the meter is not already in Setup
mode, press
.
The date format flashes in the upper
right corner.
PREPARING FOR A TEST
1. Gather Items
• CoaguChek XS Meter
• Container of test strips
• Test strip code chip
• CoaguChek lancet
TESTING
3. Match Code
Confirm that the number displayed
matches the number on the test strip
container, then press
.
If the numbers are different, first make
sure that the correct code chip is inserted.
If you are using the correct code chip
but the numbers don't match, see the
User Manual.
An hourglass appears as the meter warms
up. A flashing test strip and blood drop
appear when ready for a sample.
You have 120 seconds to apply blood to
the test strip.
6. Record Result
Record the result.
®
XS
Getting Started
2. Set Date
Press
. The year flashes.
to change the year then
.
to change the month then
to change the day then
.
The time format flashes in the upper
left corner.
2. Match Code
The code number on the test strip
container and the code chip must
match.
Each box of test strips comes with a
matching code chip. Every time you open
a new box of test strips, you must replace
the code chip.
Before continuing, review these
tips for getting a good blood drop.
Increasing the blood flow in the finger
will help you get a good drop of blood,
so keep in mind these tips:
• Warm the hand. Have the patient
hold it under his or her arm, use a
hand warmer, and/or wash with
warm water.
• Have the patient let that arm hang
by his or her side.
• Massage the finger from its base.
Use these techniques until the fingertip
has good color.
4. Collect Sample
Twist the protective cap off the
CoaguChek lancet.
Massage the finger until you see
increased color in the fingertip.
Keeping the hand down, press the
tip of the lancet firmly against the
side of the fingertip. Press the blue
trigger button.
Gently squeeze from the base of the
finger to develop a hanging drop of blood.
7. Clean Up
Place the used test strip and lancet in
an approved container.
Turn the meter off
.
If the meter is dirty, wipe it clean with a
CoaguWipe
TM
Bleach Towel.
GETTING STARTED
Follow these steps to get started using the meter:
1. Watch the CoaguChek XS System Video. It will help you get comfortable
with the CoaguChek XS Meter and the testing procedure.
2. With this CoaguChek XS System Getting Started guide by the meter, follow
the steps to perform your first test.
Then, as necessary, refer to the User Manual:
The CoaguChek XS System User Manual is a comprehensive guide to the
meter and test strips. It is designed to provide answers to your questions
about the meter's operation and use.
INSTALLING BATTERIES & SETUP
1. Open Compartment
Open the battery compartment on the
back of the meter.
Right after you insert the batteries, you'll need to set the date and time.
The date and time settings are important. Each time you run a test, the meter
compares its date with the test strip's expiration date. If the test strip is expired,
the meter displays an error message and prevents you from running a test.
Whenever you put batteries in the meter, it automatically goes to Setup mode
(where you set the date and time). You can also go to Setup mode at any time by
pressing the SET button
3. Set Time
Press
. The hour flashes.
to change the hour then
.
to change the minutes then
Turn the meter off
.
3. Insert Code Chip
Make sure the meter is off.
With the code number facing up,
insert the code chip into the code chip
slot until it snaps into place.
1. Get Ready
Take a test strip out of the container.
Close the container tightly.
Do not open a vial of test strips or touch a
test strip with wet hands or gloves. This
may damage the test strips.You have 10
minutes to use a test strip once you
remove it from the container.
5. Apply Sample
Find the target area on the test strip.
You can dose from the side or top.
See the User Manual for more
information.
Within 15 seconds of sticking the
fingertip, apply the blood to the target
area on the test strip.
Hold the blood drop to the test strip until
you hear a beep. The flashing blood drop
symbol will disappear.
Note: If during testing the meter displays an
error message, refer to the Error Messages
section of the User Manual for an explanation
and steps on how to proceed.
2. Insert Batteries
Insert 4 AAA batteries according to
the diagram inside the battery
compartment.
.
4. Check the Display
Press and hold the ON-OFF button
.
Make sure all the letters, numbers,
and symbols on the display appear
.
correctly.
Release
.
Turn the meter off
.
4. Wash Hands
Have the patient wash his or her
hands in warm, soapy water. Or, clean
the fingertip with an alcohol wipe.
Make sure the fingertip is thoroughly dry.
2. Insert Test Strip
Slide the test strip into the test strip
guide in the direction of the arrows
until it stops.
The meter turns on.
The code number of the inserted code
chip flashes on the display.
Do not add more blood to the test
strip. Do not touch the test strip.
The result appears in about 1 minute.