Aktek ALIT-TIF User Manual - Page 10

Browse online or download pdf User Manual for Measuring Instruments Aktek ALIT-TIF. Aktek ALIT-TIF 16 pages. Ultrasonic level meter

Aktek Endüstriyel Ekipman ve Enstrümantasyon Ltd. Şti
2.2.2 mounting measures
There are several methods of mounting the instrument.
 Screw mounted
Hints: Always use the nut to screw the probe.
 Flange mounted
If the maximum level to be measured falls within the blocking distance,
the transmitter must be mounted on a nozzle using an adapter flange.
Hints:
a) Select as big a nozzle diameter as possible, but keep the height as
small as possible.
b) The inner surface of the nozzle should be as smooth as possible (no
edges or welding seams).
c) The limit of nuzzle as follow:
Measuring range
4m、8m
 Bracket mounted
The mounting bracket can be used on open tanks or above conveyor
belts. The sensor is screwed into the socket provided.
Hints: Always use the nut to screw the probe.
www..aktek.com.tr
Min diameter
65mm
9
Max length
400mm