Dongguan Xin Bao Instrument Transport Simulation Vibration Table Operation Manual - Page 6
Browse online or download pdf Operation Manual for Industrial Equipment Dongguan Xin Bao Instrument Transport Simulation Vibration Table. Dongguan Xin Bao Instrument Transport Simulation Vibration Table 7 pages.
Dongguan Xin Bao Instrument Co., Ltd
IV. Environmental Requirements for Installation
1. The distance between the machine and the wall or another instrument shall be over 60cm.
2. For a stable performance of the machine, the environmental temperature shall be 15℃-30℃.
(the temperature should not change abruptly), relative humidity shall be less 85%.
3. The machine shall be installed without direct sunshine.
4. The machine shall be installed in a well-ventilated and less dusted place.
5. The machine shall be installed away with flammable and explosive materials.
6. The machine shall be installed with the power supply as near as possible. ONLY single-phase
220V AC power is accepted.
4. The machine must be reliably grounded, otherwise there is a risk of electric shock.
5. The power input terminal should be connected with the same capacity with leakage protection
of the air switch and contactor, so that when the power can be cut off immediately in emergency.
6. When the machine is running, do not touch the parts other than the control panel to prevent
abrasions or abrasions.
7. To move the machine, please be sure to turn off the power and cool it for 5 minutes.
V. Preparation before Installation
1. Please check if the power cords and ground wire are properly connected and in accordance with
the specifications.
2. The machine shall be horizontally installed to the ground.
3. Fix and clamp the specimen by adjusting the jig and place them into a balanced guardrail
device. The clamping force shall be appropriate, so as not to pinch the test samples.
VI. Operation Instruction
1. Put the specimen on the centre of the table.
2. Wretch the nut on the positioning bar.
3. There must be 1.5~1m space left around the machine.
4. Connect the power. Turn on the machine.
5. Press the key "START" for motor. (Green light is on)
6. Set the time. (00M00/min)(00H00/hour)
Eg. 05M00=Working 5 minutes (Automatically stop)
01H00=Working 1 hours (Automatically stop)
Add: 3
rd
Industrial Area, Jingshan Village, Chashan Town, Dongguan
- 3 -
Tel:0769-33365786 Fax:0769-23051695