4. 조작 방법

How to operate

4-1.
Please only use the exclusive AC code (3-tube) for the power cord. When AC
1
is not grounded, please make the connection by using the ground connector
which is located on the backside of lamp.
Please check the AC power selecting switch and then supply power to the
2
device. If the lamp is not lighted, please check the fuse and power selecting
switch.

4-2. Sensor connection

Connector is MIL-C- MS3112(14-19) connector of MIL-C-26482 Series. (contact
resistance, insulation resistance is superior)
<MS3112 14-19>
Pin No.
A
Excitation +
B
Excitation Sense +
C
Excitation –
D
Excitation Sense -
E
Signal +
F
Signal -
G
Shield
H
HB
J
D350Ω
K
D120Ω
L
R3
M
null
N
R4
P
null
R
null
S
null
T
null
U
null
V
Chassis GND
10
Use