5. Initial software setup

After mounting your cogvisAI sensor in the appropriate position and aligning it accord-
ingly, start the device by plugging in the power cable and connecting it to the power
outlet with by using the power supply unit.

Step 1: Check status

Now please check the status of your device by logging on to
https://platform.cogvis.ai. You should see the sensor with the serial number (located
on the bottom of the package) in the system.
Meaning of the sensor status on the platform:
Active: Sensor is operating.
Inactive: Operation was stopped / fall detection was paused.
Offline: Device is not connected to the internet or unplugged.
Is your sensor inactive?
-
Make sure the sensor is aligned correctly and is able to sense the floor.
-
Make sure the sensor is not covered by objects, curtains or furniture.
Is your sensor offline?
-
Check if the red light on the device is on.
-
Check if the device is powered.
-
Unplug the device and plug it back in. Wait a few minutes.
If your sensor is inactive for a longer period of time or still offline 15 minutes after
plugging it in and out, please contact our support team at +43 1 236 0580 14 or

Step 2: Automatic sensor calibration

Calibration, i.e. the adjustment of the sensor to the room structure, is carried out auto-
matically. The floor is colored yellow. If problems occur, check the position of the sen-
sor. Realign if necessary and click on "restart" on the platform. Now check the image
view.