Error Messages

Error Message
Er EP
Er Ad
Err 0
Errsp
ou ld
Gs ol
13
What to Do or Check
The setup parameters loaded in nonvolatile memory
have become corrupted.
If this problem persists, contact Doran Technical
Support for assistance.
The A/D communication is not detected. Ensure the
wire harness was not pinched during assembly.
If this problem persists, contact Doran Technical
Support for assistance.
The calibration zero is out of range.
Error is displayed after a zero calibration attempt.
Refer to the Calibration Troubleshooting section for
additional information. If this problem persists,
contact Doran Technical Support for assistance.
The span calibration is out of range.
Error is displayed after a span calibration attempt.
Refer to the Calibration Troubleshooting section for
additional information. If this problem persists,
contact Doran Technical Support for assistance.
The scale is in overload. The load on the scale
platform exceeds the scale capacity by more than
105%.
Remove excess weight from scale platform.
If problem persists, recalibrate. If this problem still
persists, contact Doran Technical Support for
assistance.
The scale is in an extreme overload condition. The
load on the scale platform exceeds the scale
capacity by more than 150%.
Immediately remove excess weight from scale
platform as damage not covered under the warranty
can occur. If problem persists, recalibrate. If this
problem still persists, contact Doran Technical
Support for assistance.