Ametek ORTEC 4002P Operating And Service Manual - Page 13
Browse online or download pdf Operating And Service Manual for Power Supply Ametek ORTEC 4002P. Ametek ORTEC 4002P 14 pages. Portable power supply
![Ametek ORTEC 4002P Operating And Service Manual](/uploads/products/server2/1164891/1164891.png)
When the dc voltage levels, as monitored at the
front panel test points, are not within the prescribed
tolerances listed in Section 6.1, each can be
adjus ted.
The
four
potentiometers are mounted on the printed wiring
board and are accessible with the left s ide panel
removed from the 4002P Portable Power Supply
module.
6.3. FACTORY REPAIR
calibrating
screwdriver
7
This instrument can be returned to the ORTEC
factory for service and repair at a nominal cost. Our
standard procedure for repair ensure the same
quality control and checkout that are used for a new
instrument. Always contact the ORTEC Customer
Service Department at (865) 482-4411, before
sending in an instrument for repair to obtain
shipping instructions and the required Return
Authorization Number. Write this number on the
address label and on the package to ensure proper
handling when the instrument reaches the factory.