D.O.R.C. MICRO DIATHERMY SYSTEM - Service Manual
After previous adjustments are performed, check the following:
- Preset position "1" =>the RF output has a value of maximum 1.5 W
If applicable, re-tune the RF output with P2.
Repeat if necessary adjustment procedure for an optimum result.
Fig. 9 - RF Output at maximum preset setting
15/17