Cole Parmer EW-36223-04 Instruction Manual - Page 8

Browse online or download pdf Instruction Manual for Controller Cole Parmer EW-36223-04. Cole Parmer EW-36223-04 20 pages. Heating controller

Cole Parmer EW-36223-04 Instruction Manual
5.

INSTALLATION

5. INSTALLATION.
5.1. Electrical safety and Installation
5.1. Electrical safety and Installation.
5.2. This equipment is designed for safe operation under the following conditions:-
5.2. This equipment is designed for safe operation under the following conditions:-
• Indoor use.
Indoor use.
• Altitude up to 2000 meters.
Altitude up to 2000 meters.
• Temperatures between 5°C and 40°C.
Temperatures between -10°C and +50°C.
• Maximum relative humidity 80% for temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C.
Maximum relative humidity 80% for temperatures up to 31°C decreasing
linearly to 50% relative humidity at 40°C.
• Mains supply voltage fluctuations up to ± 10% of the nominal voltage.
Mains supply voltage fluctuations up to ± 10% of the nominal voltage.
• Transient overvoltages typically present on the mains supply. (Overvoltage
Transient overvoltages typically present on the mains supply. (Overvoltage
category II).
category II).
• Applicable rated pollution degree 2.
Applicable rated pollution degree 2.
5.3.
This equipment must be earthed / grounded to a fixed earth / grounded mains
5.3.
This equipment must be earthed / grounded to a fixed earth / grounded mains
socket outlet. The mains supply is to be earthed / grounded in accordance with
socket outlet. The mains supply is to be earthed / grounded in accordance with
current legislation.
current legislation.
5.4. Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure
5.4. Ensure only the correct rated mains input fuses are fitted. (Where applicable ensure
the correct mains cable fuse if fitted). See Technical Specification, Section 8 of this
the correct mains cable fuse if fitted). See Technical Specification, Section 8 of this
Instruction book.
Instruction manual.
5.5. Check the voltage on the product data label of this unit. Ensure the rating conforms
5.5. Check the voltage on the product data label of this unit. Ensure the rating conforms to
to your local supply. Only connect to resistive loads = 1500Watts @ 230V~AC or
your local supply. If an 115V product is used in a 230V supply area, then a 230/115V
750Watts @ 115V~AC. Refer to Technical Specification, Section 8 of this Instruction
book.
transformer must be installed between the 230V mains supply and the product, by a
competent electrician.
5.6. This product must be connected to a mains supply source which incorporates an
RCD or GFCI device.
Only connect controller to resistive loads. Refer to section 8.
5.7. The unit is supplied with a moulded mains cord and plug set wired as follows:-
5.6. It is recommended this unit be connected to a mains supply source which incorporates
an RCD or GFCI device.
Green /
The unit is supplied with a moulded mains cord and plug set wired as follows:-
Green /Yellow or Green = Earth / Ground
Blue
Blue or White = Neutral
Brown
Brown or Black = Live / line hot.
5.8. If the controller is to be used on a retort rod, make sure it isn't positioned over any
heating apparatus.
5.7. If the controller is to be used on a retort rod, make sure it isn't positioned over any
heating apparatus.
5.9.
If using the wall mounting bracket provided, carefully mark and drill two
suitable holes using the bracket as a template. Ensure there are now cables etc
buried in the wall. Plug the holes with raw plugs and fasten the bracket to the wall in
the correct orientation using the screws provided. Place the controller over the
bracket.
Page 8 of 24
or
Yellow
or
or
5. INSTALLATION.
5.1. Electrical safety and Installation.
5.2. This equipment is designed for safe operation under the follo
Indoor use.
Altitude up to 2000 meters.
Temperatures between -10°C and +50°C.
Maximum relative humidity 80% for temperatures up to 31°
linearly to 50% relative humidity at 40°C.
Mains supply voltage fluctuations up to ± 10% of the nomin
Transient overvoltages typically present on the mains supp
category II).
Green
=
Earth / Ground
Applicable rated pollution degree 2.
White
=
Neutral
5.3.
This equipment must be earthed / grounded to a fixed earth /
socket outlet. The mains supply is to be earthed / grounded in acc
Black
=
Live / line hot.
current legislation.
5.4. Ensure only the correct rated mains input fuses are fitted. (Where
the correct mains cable fuse if fitted). See Technical Specification,
Instruction book.
5.5. Check the voltage on the product data label of this unit. Ensure the
to your local supply. Only connect to resistive loads = 1500Watts @
750Watts @ 115V~AC. Refer to Technical Specification, Section 8
book.
5.6. This product must be connected to a mains supply source which in
RCD or GFCI device.
5.7. The unit is supplied with a moulded mains cord and plug set wired
8
M6493 Issue 10.3
or
Green /
Yellow
Green
=
Earth / Gr