INTENDED USE AND GUIDELINES
Intended Use
1. EGA-500 Irrigation Pump – Irrigation Delivery system
2. EGA-501 CO2 Insuf ator – Medical CO2 delivery system
PRODUCT FAMILIARIZATION
EGA-500 Irrigation Pump Components and Their Functions
1. Irrigation Pump Unit - facilitates irrigation delivery
2. Pump Unit base – connects to irrigation unit and securely attaches water bottle holder
3. Water bottle holder/warmer – plugs to rear of irrigation unit to incorporate water warming feature
4. Tool kit – includes screws and Allan wrenches to attach water bottle holder to the base and the base to the unit.
5. Peristaltic pump head – attaches to pump for irrigation delivery
6. Power cord
Consumables / Accessories
1. 100130 EndoGator® irrigation tubing. The 200230 irrigation tubing is not compatible with this unit.
2. 1006## and 1006## EndoGator® Hybrid tube sets. The 1006## and 1006## are not compatible with this unit.
3. 1006## and 1006## tubing kits. The 1006## and 1006## tubing kits (with 200230 tube sets) are not compatible with this unit.
EGA-501 CO2 Insuf ator Components and their Functions
1. CO2 Insuf ator Unit - facilitates CO2 delivery
2. Insuf ator Unit base – fastens to insuf ator unit and securely attaches water bottle holder
3. Water bottle holder/warmer – plugs to rear of insuf ator unit to incorporate water warming feature
4. Tool kit – includes 2 sets of screws and 2 sets of Allan wrenches to attach water bottle holder to the base and the base to the unit.
5. Power cord
IF CO2 SOURCE IS A TANK:
6. High Pressure Tank hose (thin with twin connections) to attach to the Insuf ator unit and the source
7. Tank yoke to secure hose and maintain CO2 ow
IF CO2 SOURCE IS FROM A WALL SUPPLY:
8. Low pressure tank hose (thicker rubber tubing) with brass end to connect to the insufflator and a silver end to connect to the wall.
Consumables / Accessories
1. All CO2 Endo SmartCap® tube sets.
2. 100551 CO2 source tubing. This is needed for use with the CO2 Endo SmartCap® products, but not for the Hybrid tube sets.
3. All EndoGator® Hybrid tube sets. However, note that the choice of EndoGator® Hybrid must take into account the irrigation pump.
SYSTEM FUNCTIONS
EGA-500 Irrigation Pump
Irrigation Pump Front Panel:
1. On/Off Button functions properly and actuates to turn pump on/off.
2. H2O button remains green to indicate proper functionality. It will turn blue if the heater is unplugged.
3. Explanations of the priming function. Button remains lit while priming only.
4. Flow rate knob should move freely without hesitation.
Procedure
1. Adjust ow rate if needed
2. If the bottle needs to be replaced, follow instructions on tube set IFU regarding use of sterile technique.
3. If the CO2 ow is stopped for any reason, turn on the air pump to maintain pressure in the air/water system any time the
endoscope is in the patient.