Manufacturer's Declarations
Guidance and Manufacturer's Declaration - Emissions
The display is intended for use in the electromagnetic environment specified below.
The customer or user of the display should ensure that it is used in such an environ-
ment.
IEC 60601
Immunity Test
Test Level
Conducted
3 Vrms
RF
150 kHz to 80
IEC 61000-4-6
MHz
3 V/m
Radiated RF
80 MHz to 2.5
IEC 61000-4-3
GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people
15
(continued)
Compliance
Level
Environment - Guidance
Portable and mobile communications
equipment should be separated from
Display (195) by no less than the
distances calculated/listed below:
N/A
D=(3.5/V1)(Sqrt P)
3 V/m
D=(3.5/E1)(Sqrt P) 80 to 800 MHz
D=(7/EI)(Sqrt P) 800 MHz to 2.5 GHz
where P is the max power in watts and D
is the recommended separation distance
in meters.
FIeld strengths from fixed transmitters,
as determined by an electromagnetic site
survey, should be less than the compli-
ance level in each frequency rangeb.
Interference may occur in the vicinity
of equipment containing a transmitter
symbol:
Electromagnetic
exerspy display