The Endevco Model 22 is a microminiature, piezoelectric accelerometer. Because of its extremely small
size and physical construction, certain special precautions must be observed when mounting and
removing the accelerometer to avoid damaging the unit.
Accessories, Adhesive, Solvent and Tools
For use at room temperature:
* Petro wax or bees wax
For use up to 60°C:
* All-temperature glue stick and hot-melt glue gun
* Solvent: Follow the recommendations from glue stick manufacturers
For use up to 125°C:
* Adhesive: Cyanoacrylate (Loctite Super Bonder 430 Adhesive is recommended)
* Solvent: Nitromethane base (Loctite X-NMS Solvent 768 is recommended)
Alternatives: Acetone, Chlorothene (Trichloroethane) or equivalent hydrocarbon solvent
* Cotton swabs or lint-free tissue
* Toothpicks or equivalent applicators
Endevco P/N 31849 Adhesive Kit contains most of the required items for mounting transducer.
Contact the factory for details.
CAUTION:
Observe precautions recommended by manufacturer of adhesive or solvent.
Special Precautions
The following precautions should be observed when performing the accelerometer mounting and removal
procedures given below.
WARNING:
Do not break cement bond by striking accelerometer or by using any tool that
makes a metal-to-metal contact with the accelerometer. Always break the bond
with the supplied tool after the cement has been softened, by twisting or torquing
the accelerometer, stressing the bond in shear.
CAUTION:
Exercise care to prevent getting adhesive or solvent on the accelerometer
connector.
Mounting Procedure
1. Ensure that mating surfaces are smooth and flat.
2. Remove all traces of grease, oil and oily residue from mating surfaces using cotton swabs and
solvent.
3. Apply a tiny drop of Cyanoacrylate, approximately 1/5 the diameter of the base mounting surface
(side opposite gold lid).
4. Spread the cement with toothpick to form thin, uniform layer covering area of contact, and
immediately position accelerometer on mating surface.
5. Press down and hold accelerometer firmly in place for 30 seconds while cement sets. Allow
cement to set for at least several minutes for maximum strength.
6. If the accelerometer is to operate at low frequencies and/or large acceleration levels, the
accelerometer cable must be secured to the vibrating structure as close as possible to the
accelerometer to avoid cable damage.