Amphenol Tempsonics Level Plus LLE Operation Manual - Page 12
Browse online or download pdf Operation Manual for Transmitter Amphenol Tempsonics Level Plus LLE. Amphenol Tempsonics Level Plus LLE 20 pages. Magnetostrictive liquid level transmitters with temposonics technology
Level Plus
LLE
®
Operation Manual
Electrical connection
Connection type
Operating voltage
Ripple
Current consumption
Dielectric strength
Polarity protection
Overvoltage protection
5.7 Technical drawing
M12 Connector
13
(0.5)
Sensor electronics housing
Controlling design dimensions are in millimeters and measurements in ( ) are in inches
Fig. 2: Technical drawing LLE
Analog:
M12 male connector (5 pin)
SSI:
M12 male connector (8 pin)
IO-Link:
M12 male connector (4 pin)
CAN:
M12 male connector (5 pin)
+24 VDC (−15 / +20 %); UL recognition requires an approved power supply with energy
limitation (UL 61010-1), or Class 2 rating according to the National Electrical Code (USA) /
Canadian Electrical Code
≤ 0.28 V
PP
Analog: 50...140 mA, SSI, CANbus & IO-Link: 60 mA
500 VDC (DC ground to machine ground)
Up to −30 VDC
Up to 36 VDC
Process connection
NPT (3/4", 1", 1.5", 2", 3", 4") or
RF Flange (2", 3", 4", 5", 6")
57
51
(2.2)
(2.0)
Null zone
Product float
Sensor rod
Type: 7 (0.28) OD
Type: 10 (0.39) OD
50...2500
(2...100)
Measurement range
156...2646
(6.2...104.2)
Order length
I
12
I
Stop collar
Interface float
63.3
(2.5)
Inactive zone