Foldawheel PW-4 4Q Series User Manual - Page 16
Browse online or download pdf User Manual for Wheelchair Foldawheel PW-4 4Q Series. Foldawheel PW-4 4Q Series 20 pages.

General Guidelines
Avoid knocking or bumping the controller, especially the joystick. Avoid prolonged
exposure of your chair to extreme conditions, such as heat, cold, or moisture.
Keep the controller clean
Check all connectors to ensure that they are all tight and secured properly. WARN-
ING! Even though the electric wheelchair has passed the necessary testing require-
ments for ingress of liquids, you should keep electrical connections away from
sources of dampness, including direct exposure to water or bodily fluids and inconti-
nence. Check electrical components frequently for signs of corrosion and replace as
necessary.
Daily Checks
With the controller turned off, check the joystick. Make sure it is not bent or damaged
and that It returns to the neutral position when you release it, Check the rubber boot
around the base of the joystick for damage. Visually inspect the boot. Do not handle
or try to repair it, See an authorized provider if there is a problem.
Visually inspect the controller cable. Make sure that it is not frayed, cut, or has any
wires exposed. See an authorized provider if there is a problem.
Check for flat spots on solid tires. Flat spots could adversely affect stability.
Weekly Checks
Disconnect and inspect the controller from the power base. If there is any corrosion,
contact and authorized provider if necessary.
Ensure that all parts of the controller system are securely fastened to your electric
wheelchair.
Calibrate the joystick if a noticeable difference in performance is detected or if the
joystick does not operate properly.
To calibrate the joystick, power off the unit, place the joystick in the neutral position,
and power the unit back on. .If a problem still exists with your joystick's performance,
contact an authorized provider.
Check the brakes. This test should be carried out on a level surface with at least 3
feet (1 meter) of clearance around your Z-Chair.
Check the brakes.
1. Turn on the controller and turn down the speed level of your electric wheelchair.
2. After one second, check the battery condition meter. Make sure that it remains on.
3. Slowly push the joystick forward until you hear the electric brakes click.
Immediately release the joystick. You must be able to hear each electrical brake
operating within a few seconds of joystick movement. Repeat this test three times,
pushing the joystick backwards, then left, and then right.
15