Amprobe ACD-30P User Manual - Page 16

Browse online or download pdf User Manual for Measuring Instruments Amprobe ACD-30P. Amprobe ACD-30P 19 pages. 1000a clamp-on power meters

Accessories : Test leads (MTL-90B), batteries installed, user's manual, soft carrying
pouch, & TPK-59 banana plug type-K thermocouple (model ACD-31P & ACD-41PQ
only)
Optional accessories : RS-232 KIT2 interface kit
Electrical Specifications
Accuracy is ±(% reading digits + number of digits) or otherwise specified, at 23
&
less than 75% R.H.
True RMS (all models) ACV & ACA clamp-on accuracies are specified from 0% to 100%
of range or otherwise specified. Maximum Crest Factor is as specified below, and with
frequency spectrums, besides fundamentals, fall within the meter specified AC
bandwidth for non-sinusoidal waveforms. Fundamentals are specified at 50Hz and
60Hz.
AC Voltage
RANGE
50Hz / 60Hz
600.0V
45Hz ~ 500Hz
600.0V
500Hz ~ 3.1kHz
600.0V
CMRR : >60dB @ DC to 60Hz, Rs=1kΩ
Input Impedance: 2MΩ, 30pF nominal
Crest Factor:
< 2.3 : 1 at full scale & < 4.6 : 1 at half scale
ACV AutoVA
Threshold: 30VAC (40Hz ~ 500Hz only) nominal
TM
DC Voltage
RANGE
600.0V
NMRR : >50dB @ 50/60Hz
CMRR : >120dB @ DC, 50/60Hz, Rs=1kΩ
Input Impedance: 2MΩ, 30pF nominal
DCV AutoVA
Threshold: 2.4VDC nominal
TM
Accuracy
0.5% + 5d
1.5% + 5d
2.5% + 5d
Accuracy
0.5% + 5d
15
C ±5
C
o
o