2.1 Product Installation
1. Please do not lay the product on the floor or subject it to shock. It can cause damage inside
the product.
2. Please install the product in a hard, flat, and stable place.
3. Please do not install the product in a place where it is exposed to direct sunlight, or in the
dangerous places.
4. To minimize the risk of short circuits, do not install the product (in the) places where it will
be subjected to the inflow of moisture or organic solvents on the inside of the body.
5. This product was manufactured for normal operation at a rated voltage, so please check the
voltage status before installing the product.
6. When installing and using the product, please certainly use the power cord provided with
the products.
7. Please be careful so that moisture, organic solvent, dust, corrosive and gas, etc. do not
enter into this product.
8. Please do not expose this product in the place where there is direct heat sources.
* Caution
The warnings sticker is affixed to the product. Please read.
-
Example) WUC-D06H
2. Installation
DAIHAN Scientific, Analog Ultrasonic Cleaners " WUC-A "
* WUC-A01H
Do not make water level below 3 c m
-
Do not operate continuously more than 5 minutes
-
Keep 5 minutes break time when after 5 minutes run
-
continuously.
* WUC-A02H/A03H
Do not make water level below 6 c m
-
Do not operate continuously more than 30 minutes
-
Keep 10 minutes break time when after 30 minutes
-
run continuously.
* WUC-A06H/A10H/A22H
Do not make water level below 10 c m
-
Do not operate continuously more than 30 minutes
-
Keep 10 minutes break time when after 30 minutes
-
run continuously.
* Common Point
Turn off the heater power if the unit is not going to
-
be used.
When make water level below standard, it may cause
-
damage of heater.
Do not use acid detergent solution(Chlorine line).
-
6/36