Amprobe AM-105 TRMS User Manual - Page 11

Browse online or download pdf User Manual for Multimeter Amprobe AM-105 TRMS. Amprobe AM-105 TRMS 15 pages.

Amprobe AM-105 TRMS User Manual
10
Battery and Fuse replacement
Battery use:
Standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Protection-reinforced CE Version fuses:
Fuse (FS1) for µAmA current input: 0.63A/500V, IR 200kA, F fuse (FA-6x32/.63);
Fuse (FS2) for A current input: 12.5A/500V, IR 20kA, F fuse (FA-12.5A/500V);
Battery replacement:
Loosen the 2 screws from the battery access door of the case bottom. Lift the battery
access door and thus the battery compartment up. Replace the battery. Re-fasten the
screws.
Fuse replacement:
Loosen the 4 screws from the case bottom. Lift the end of the case bottom nearest the
input jacks until it unsnaps from the case top. Replace the blown fuse(s) and/or the
battery. Replace the case bottom, and ensure that all the gaskets are properly seated
and the two snaps on the case top (near the LCD side) are engaged. Re-fasten the
screws.