Amptek X-123 User Manual - Page 9

Browse online or download pdf User Manual for Measuring Instruments Amptek X-123. Amptek X-123 12 pages.
Also for Amptek X-123: Quick Start Manual (4 pages), User Manual (18 pages)

3.3

Connectors

Power
Power Jack on X-123SDD: Hirose MQ172-3PA(55) [obsoleted in 2016, replaced by MQ172-3PA(30)]..
Mating Plug: Hirose MQ172-3SA-CV(50) [obsoleted in 2016, replaced by MQ172-3SA-CV(30)]
[The replacements are form/fit/function compatible.]
Auxiliary
2
16-pin
mm spacing
Mates with cable assembly (Samtec P/N) TCSD-08-S(D)-xx.xx-01-F-N.
Pin #
Name
1
SCA1
3
SCA3
5
SCA5
7
SCA7
9
AUX_IN_1
11
AUX_IN_2
13
IO2
15
GND
X-123 User Manual
Pin #
Name
1
VIN
2
GND
3
Do Not Connect
Pin #
Name
2
SCA2
4
SCA4
6
SCA6
8
SCA8
10
AUX_OUT_1
12
AUX_OUT_2
14
IO3
16
GND
9
Pin 1
Pin 2