DPM TT 570 S Operating Manual - Page 7

Browse online or download pdf Operating Manual for Measuring Instruments DPM TT 570 S. DPM TT 570 S 16 pages. Micromanometer

 
The User Menu is used to input user values and change default settings.
To access the User Menu, switch on the instrument while holding down ENTER. 
 
Page 05 
Set Clock: 
Auto Zero Time: 
 
Page 06 
Switch Off Period: 
RS232 Period: 
Note: 
Back Light Period: 
Temp / Pressure: 
 
Page 07 
Select Date Format: 
 
Selecting from a list: 
Entering a value: 
 
Set Clock 
Auto Zero Time 
Switch Off Period 
RS 232 Result Period 
Back Light Period 
Temp / Pressure 
Select Date Format 
Exit 
 
Set Clock 
Auto Zero Time 
Switch Off Period 
RS 232 Result Period 
Back Light Period 
Temp / Pressure 
Select Date Format 
Exit 
 
USER MENU
To set the clock to local time and input the date in the chosen format.
This is factory set at 60 seconds.  
(for which the instrument specifications are based) 
Changing the default setting will override this.   
Inputting 0 seconds will switch off the auto zero function.   
A manual zeroing facility is available see Keypad Controls page 02. 
 
 
To save battery the instrument will switch off after 10 minutes providing that no 
keys are pressed.  This can be overridden by changing the default setting. 
To print out raw data as measurements are being made, either to a thermal 
printer or to a PC via the Hyper terminal. 
DP Measurement no longer supply thermal printers.
To alter the backlight time. Use  and  to switch on the back light.
To change the air stream temperature and barometric pressure. 
Air stream temperature set at 16 °C and Barometric pressure set at 1000 mbar. 
 
 
Set the date format to European or USA, the temperature units to °C or °F and 
the barometric pressure units to mb or inHg. 
 
Use  and  to scroll.  Press ENTER to select.
Use   and    to set each digit or value then press ENTER.  
Press ENTER
Press ENTER
TT Series LR Manual 
Select Date and Time
Minutes?
16
Hours?
14
Date?
26
Month?
10
Year?
10
User Menu
AutoZero Period
060 Seconds
Max Setting:
Min Setting:
To switch off input:
User Menu
Page 05
 
 
 
 
 
 
 
 
 
 
 
 
 
600  
010  
000