Amrobe 34XR-A User Manual - Page 11
Browse online or download pdf User Manual for Multimeter Amrobe 34XR-A. Amrobe 34XR-A 19 pages. Professional digital multimeter
Approvals:
>
LISTED
950Z
Safety: Conforms to UL1244; EN61010-1:
Cat II - 1000V / Cat III - 600V; Class 2,
Pollution degree II.
EMC: Conforms to EN61326-1, criteria B
This product complies with requirements of
the following European Community
Directives: 89/ 336/ EEC (Electromagnetic
Compatibility) and 73/ 23/ EEC (Low
Voltage) as amended by 93/ 68/ EEC (CE
Marking). However, electrical noise or
intense electromagnetic fields in the vicinity
of the equipment may disturb the
measurement circuit. Measuring
instruments will also respond to unwanted
signals that may be present within the
measurement circuit. Users should exercise
care and take appropriate precautions to
avoid misleading results when making
measurements in the presence of electronic
interference.
Electrical Specifications
(Accuracy at 23 °C ±5 °C, <75 % relative
humidity)
DC VOLTS
Ranges: 400mV, 4V, 40V, 400V, 1000V
Resolution: 100 µV in 400mV range
Accuracy: ±(0.5 % rdg + 1 dgt)
Input impedance:
Ω
400mV: >100 M
; 4V: 10 M
Ω
1000V: 9.1 M
Overload protection: 1000 V dc /
750 Vac rms
AC VOLTS true rms
Ranges: 400m, 4V, 40V, 400V, 750V
Resolution: 100 µV
Accuracy:
±(1.2 % rdg +8 dgts) 45 Hz to 100 Hz on
400mV range
±(1.2 % rdg + 8 dgts) 45 Hz to 500 Hz
±(2.0 % rdg +8 dgts) 500 Hz to 2 kHz
±(2.0 % rdg + 8 dgts) 45 Hz to 1 kHz on
750 V range
Crest Factor: ≤ 3
P
Ω
; 40V to
(45Hz - 2kHz)
Input impedance: 400mV: >100 M
Ω
4V: 10 M
; 40V to 1000V: 9.1 M
AC coupled true rms specified from
5 % to 100 % of range
Overload protection: 1000 V dc or
750 V ac rms
DC CURRENT
Ranges: 400µA, 4000µA, 40mA, 300mA,
10A
Resolution: 0.1µA
Accuracy: ±(1.0 % rdg + 1 dgt) on 400µA to
300mA ranges
±(2.0 % rdg + 3 dgts) on 10A range
Burden voltage:
µ
400
A Range:
4 mA Range:
40 mA Range:
300 mA:
10A:
Input protection: 0.315A/1000V fast blow
ceramic fuse 6.3×32mm on µA/mA input
10A/1000V fast blow ceramic fuse
10×38mm on 10A input
10A input: 10 A for 5 minutes maximum
followed by a 10 minute cooling period
AC CURRENT true rms (45Hz - 1kHz)
Ranges: 400µA, 4000µA, 40mA, 300mA,
10A
Resolution: 0.1 µA
Accuracy: ±(1.5 % rdg + 8 dgts) on 400µA
to 300mA ranges
±(2.5 % rdg + 10 dgts) on 10A range
Crest Factor: ≤ 3
Burden Voltage: See DC Current
Input protection: 0.315A/1000V fast blow
ceramic fuse 6.3×32mm on µA/mA input
10A/1000V fast blow ceramic fuse
10×38mm on 10A input
10A input: 10 A for 4 minutes maximum
followed by a 12 minute cooling period
10
Ω
;
Ω
µ
1 mV/ 1
A
500 mV/ 1 mA
10 mV/ 1 mA
8 mV/ 1 mA
40 mV/ 1 A