AmScope PM200 Series User Manual - Page 15
Browse online or download pdf User Manual for Microscope AmScope PM200 Series. AmScope PM200 Series 16 pages.
PM200 Series
Common Issues (Optical)
Symptom
Cause
Stains, dust, or dirt has accumulated on the speci-
Obstructions are
observed in the field
men
of view
Stains, dust, or dirt have accumulated on the
lenses
Unclear Image
Stain or dust has accumulated on the lens in the
inlet of the head
The microscope head is not in the right position
to be focused
One side of the field
The specimen is not fixed
of view is dark or the
image moves while
focusing
The field of view is not
The light source is not present
bright enough
The light source is too dim
Stains, dust, or dirt has accumulated on the con-
denser, objective, eyepieces, or base lens
Remedy
OPTICAL ISSUES
Clean the slide or use a new specimen if sample is
destroyed
Clean the objective and eyepiece lenses
Clean the lens with lens cleaner or a nonabrasive
lint free cloth, as well as spray with compressed air
Adjust the height of the microscope with the focus-
ing knob on the focusing rack until image appears
in focus. Note: Higher magnification will require
the head to move closer, while lower will require
more distance between the sample and the lens.
Secure the slide to the stage with clips or another
manner so that the sample does not move
A light source must be used with this model micro-
scope to view a sample (ring light, built in illumina-
tor, or gooseneck illuminator)
The light source used is not bright enough, or is set
to too low of an intensity setting. Adjust setting to a
higher or change light sources
Thoroughly clean tall lenses with lens cleaner or a
lint free nonabrasive cloth
15