Congratulations on acquiring a new MACH LED 6MC operating
light.
Please read these Instructions for use very carefully.
1. Instructions for safe use
1.1
Intended user
The Mach LED 6MC is a Class I medical product and may only be
operated by trained medical staff.
1.2
Information and obligation of the user to check the product
Pay attention to the instructions for use when handling the lamp.
These Instructions for use are part of the product and must therefore
be stored in a place close to the product in order for the safety in-
structions and important information to be consulted at any time.
Make sure that the lamp is in satisfactory working order before every
use. If there is obvious damage, unusual operating conditions, etc.,
the lamp must not be used.
1.3
Availability of the instructions for use
These instructions for use and a detailed handbook with detailed in-
formation on how to deal with a fault, a list of accessoires and further
tips on how to use the light can be found online at the following link:
https://dr-mach.de/login/mach-led-6mc.html
1.4
Intended use / contra-indications
The Mach LED 6MC operating light is designed to illuminate an op-
erating site in medical facilities (e.g. in a laboratory, in hospitals or
doctor's practice) with focussed, low-glare, shadow-free light. It ena-
bles the user to perform a diagnosis or carry out medical interven-
tions. The Mach LED 6MC light is an operating light that is not fail-
safe when used as a single light. It is not intended for use in areas
where explosions are likely although it is permissible to use it in the
vicinity of HF surgical equipment.
Permanent illumination of the open human eye should be avoided
when illuminating the face area.
59361001
1.5
Technical data
Class of pro-
I
tection
IP protection
IP 54 (light body without camera preparation)
class
IP 53 (light body with camera preparation)
Input voltage
100-240 V AC, 50/60 Hz
(power supply)
Input voltage
24-30 V DC
(light body)
Power con-
Standard
sumption /
140.000 Lux
Current
66 W / 2,7 A max.
(without KV)
89 W / 3,3 A max.
(without KV, mit S)
65 W / 2,7 A max.
(withKV, without S)
75 W / 3,1 A max.
(with KV, with S)
75 W / 3,1 A max.
(with camera, without
S)
84 W / 3,5 A max.
(with Kamera, with
S)
Operating time
Continuous operation possible
(1)
Expected life
10 years
KV: camera preparation; S: shade management
1
At the end of the expected (designed) service life, the lamp must be serviced
more frequently for safe operation (see the manual for details).
1.6
Lighting technical data
Mach LED
6MC
Central light intensity
140,000 lux
(distance 1 m)
(optional
160,000 lux)
Light field diameter d10
180 mm
Light field diameter d50
103 mm
Residual light intensity
50%
(one shadower)
b
Residual light intensity
45%
c
(two shadowers)
58%
Residual light intensity
100%
(normed tube)
5 / 70
Residual light intensity
(normed tube, one shad-
ower)
Residual light intensity
(normed tube, two shad-
owers)
Illumination depth (20%)
Illumination depth (60%)
Radiation intensity in the
Optional
field
160.000 Lux
(distance 1 m)
75 W / 3,1 A max.
Max. radiation intensity
(without KV)
in the field
94 W / 3,9 A max.
(distance 0.69 m)
(without KV, mit S)
a
KV refers to the model with camera preparation, the lighting specifications
78 W / 3,2 A max.
have been determined for a light with a built-in camera.
(with KV, without
b
Without shade management system
S)
c
With shade management system (optional)
87 W / 3,6 A max.
d
Mach LED 6MC with 160,000 lux (optional)
(with KV, with S)
87 W / 3,6 A max.
For a complete overview of technical and lighting specifications, refer
(with camera, with-
to the manual.
out S)
95 W / 3,9 A max.
1.7
Installation/Maintenance/Repair
(with Kamera, with
S)
The lamp may only be installed, maintained or repaired by the man-
ufacturer or by specially trained staff. Maintenance must be carried
out at least every two years
1.8
Environmental conditions for operation
Ambient temperature:
Relative air humidity:
Air pressure.
1.9
Reporting obligations
Mach LED
a
6MC KV
Every serious incident which has occurred in connection with the
140,000 lux
product must be reported to the manufacturer and the competent au-
(optional
thority.
160,000 lux)
180 mm
2. Images on the device
103 mm
50%
b
41%
c
53%
100 %
Mach LED 6MC
50%
50 %
b
b
45%
41 %
c
c
58%
53 %
2000 mm
1680 mm
1050 mm
1010 mm
2
2
518 W/m
500 W/m
2 d
2 d
558 W/m
567 W/m
2
2
723 W/m
673 W/m
2 d
2 d
757 W/m
763 W/m
+5°C to +40°C
30% to 75% RH
700h Pat o 1060h Pa
This symbol indicates that the instructions
for use must be followed.
V5