Dr. Mach MACH 120 Series Instructions For Use Manual - Page 17

Browse online or download pdf Instructions For Use Manual for Lighting Equipment Dr. Mach MACH 120 Series. Dr. Mach MACH 120 Series 19 pages. Ceiling, wall, new-type stand model
Also for Dr. Mach MACH 120 Series: Mounting Instructions, Directions For Use (19 pages)

Dr. Mach MACH 120 Series Instructions For Use Manual
MACH 120F/120
Table 201 – Guidance and manufacturer´s declaration – electromagnetic emission – for all EQUIPMENT AND SYSTEMS (see
6.8.3.201 a) 3))
1
The MACH 120 / 120 F is intended for use in the electromagnetic environment specified below. The
2
customer or the user of the MACH 120 / 120 F should assure that it is used in such an environment.
3
Emissions test
Harmonic emissions
7
IEC 61000-3-2
Voltage fluctuations /
8
flicker emissions
IEC 61000-3-3
RF emissions
12
CISPR 15
Table 202 – Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS (see
6.8.3.201 a) 6))
The MACH 120 / 120 F is intended for use in the electromagnetic environment specified below. The customer
or the user of the MACH 120 / 120 F should assure that it is used in such an environment.
Immunity test
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient / burst
IEC 61000-4-4
± 1 kV for input/output lines
± 1 kV differential
Surge
mode
IEC 61000-4-5
± 2 kV common mode
< 5 % U
(>95 % dip in U
for 0,5 cycle
40 % U
Voltage dips, short
(60 % dip in U
interruptions and
for 5 cycles
voltage variations
on power supply
70 % U
input lines
(30 % dip in U
IEC 61000-4-11
for 25 cycles
< 5 % U
(>95 % dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE
U
is the a. c. mains voltage prior to application of the test level.
T
59080001
Ceiling, Wall, Stand model

9. Electromagnetic compatibility

Guidance and manufacturer´s declaration – electromagnetic emission
Compliance
Class C
Complies
Complies
Guidance and manufacturer´s declaration – electromagnetic immunity
IEC 60601
test level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
not applicable
± 1 kV differential
mode
± 2 kV common mode
< 5 % U
T
)
(>95 % dip in U
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
T
for 5 cycles
70 % U
T
)
(30 % dip in U
T
for 25 cycles
< 5 % U
T
)
(>95 % dip in U
T
for 5 sec
30 A/m
Electromagnetic environment - guidance
The MACH 120 / 120 F is suitable for use in all
establishments, including domestic establishments and those directly con-
nected to the public low-voltage power supply network that supplies buildings
used for domestic purposes.
The MACH 120 / 120 F is not suitable for interconnection
with other equipment.
Compliance level
T
)
T
T
)
T
T
)
T
T
)
T
Edition 08
Dr. Mach
Lamps and Engineering
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital environment. If
the user of the
MACH 120 / 120 F requires
continued operation during power
mains interruptions, it is recommended
that the MACH 120 / 120 F be powered from
an uninterruptible power supply or a battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
06.02.2010 / Bak Page 17/19