Immunity Test
Electrostatic
6 kV contact
discharge (ESD)
8 kV air
IEC 61000-4-2
Electrical fast
2 kV for power
transient/burst
supply lines
IEC 61000-4-4
1 kV for input/
output lines
Surge
1 kV line(s) to
IEC 61000-4-5
line(s)
2 kV line(s) to
earth
Interruptions
<5 % UT
and voltage
(>95 % dip in UT)
variations on
for 0,5 cycle
power supply
input lines
40 % UT
(60 % dip in UT)
IEC 61000-4-11
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
LifeTemp
IEC 60601 Test
Compliance
Level
6 kV
contact
8 kV air
Not
applicable
Not
applicable
Not
applicable
3 A/m
Radient
®
Electromagnetic Environment –
Level
Floors should be wood, concrete
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
TH8xyz series requires continued
operation during power mains
interruptions, it is recommended
that the TH8xyz series be powered
from an uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
Non-Contact Thermometer
®
Guidance