3.2.3 Time Settings

1.
Tap on the time display button for each cycle to change the time. Enter the temperature using
the key pad on the screen.
2.
Tap

3.2.4 Cycle Settings

1.
Tap on the purple display button
displays the number of cycles.
2.
Tap on 'CYCLES' to enter the number of cycles you want to run.
3.
Tap on 'GO TO' to enter the step number you want the cycle to repeat every time, after the
last step of the cycle.
Add or delete a step:
1.
Add a step—tap to select the step after which a
new step must be inserted.
2.
Delete a step—tap to select the step which must
be deleted.
Atila Powergene Real-Time PCR Systems User Manual
Example: to set the time to 2 minutes, Tap 'Min' on the screen and enter the number 2
using the key pad on the screen. Press 'DONE' to save.
Example 2: to set time to 1 minutes 30 seconds, tap on 'Min' on the screen and enter the
number 1. Tap on 'Sec' and enter the number 30 using the key pad on the screen. Press
'Done' to save.
to clear the entered numerical digits.
Example: To set 40 cycles consisting of step 2 and 3, tap on the purple button that
displays number of cycles. Tap on CYCLES and enter 39. Tap on 'GO TO', and enter 2
(Step number).
Note: The cycle count in the system starts at 0. Thus, 0 to 1 is considered as 1 cycles.
Thus, always enter a number less than the desired number of cycles. Tap on 'GO TO', and
enter 2.
Tap on
to add a new step.
Tap on
to delete an existing step.
under the CYCLING stage. The number on the button