caron 7001-50 Quick Start Manual - Page 5

Browse online or download pdf Quick Start Manual for Laboratory Equipment caron 7001-50. caron 7001-50 6 pages.

caron 7001-50 Quick Start Manual
6. Center new air filter with condenser edges as shown below and press into place
with cardboard side towards condenser.
filter it cannot be adjusted.
7. Lower control panel bezel by disengaging support arms on both sides.
Verify wiring harnesses are clear of control panel bezel.
8. Secure control panel bezel with retained (2) screws using Phillips screwdriver.
Note: Silver screw fastens left side of control panel bezel and black screw fastens
right side.
9. Continue to instructions for replacing additional accessory preventative
maintenance items if applicable or plug in power cord and turn on chamber.
10. Reset air filter alert by pressing Settings button on right side of Home screen as
shown below.
INST-PM-7001-50
Note: Once adhesive has touched air
Air Filter
Rev B 04/03/2018
Note:
Settings Button