Andon KD-926 Operation Manual - Page 13

Browse online or download pdf Operation Manual for Blood Pressure Monitor Andon KD-926. Andon KD-926 19 pages.

DN:KD-926-SMSY01 V1.0
Note: It is not intended to provide a basis of any type of rush toward emergency
conditions/diagnosis based on the color scheme and that the color scheme is meant only to
discriminate between the different levels of blood pressure.

12. TECHNICAL ALARM DESCRIPTION

The monitor will show 'HI' or 'Lo' as technical alarm on LCD with no delay if the determined
blood pressure (systolic or diastolic) is outside the rated range specified in part
SPECIFICACIONS. In this case, you should consult a physician or check if your operation
violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and cannot be
adjusted or inactivated. This alarm condition is assigned as low priority according to IEC
60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on LCD will
disappear automatically after about 8 seconds.

13. TROUBLESHOOTING (1)

PROBLEM
LCD Display shows
abnormal result
POSSIBLE CAUSE
The cuff position was not correct or
it was not properly tightened
Body posture was not correct during
testing
Speaking, arm or body movement,
angry, excited or nervous during
testing
Irregular heartbeat (arrhythmia)
12/16
Operation Guide
SOLUTION
Apply the cuff correctly and try
again.
Review the "BODY POSTURE
DURING
MEASUREMENT"
sections of the instructions and
re-test.
Re-test when calm and without
speaking or moving during the
test.
It is inappropriate for people with
serious arrhythmia to use this
Electronic Sphygmomanometer.