DN:KD-926-SMSY01 V1.0
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
portable and mobile RF communications equipment and the KD-926
The KD-926 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the KD-926 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the KD-926 as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output
power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be determined using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Table 4
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
d
1
2 .
P
0.12
0.38
1.2
3.8
12
17/16
Operation Guide
800 MHz to 2,5 GHz
d
2
3 .
P
0.23
0.73
2.3
7.3
23