8. TRANSFUSION :
Transfusion : for residential practice only, with classic transfusion lines by
hanging the Droper Field to a pole by the mean of the hanging ring on the
Droper Field rear side. The drip chamber MUST stay in a vertical position, with
the drip chamber end side down
= the line is purged of any air and the drip chamber is half filled.
1.
Close the transfusion line roller clamp
2.
Insert the spike into the blood bag
3. Place the blood bag in the horizontally arranged
Droper Field
4. Close the Droper Field
5. Hang the Droper Field to a pole
6. Fill the drip chamber halfway by pinching it
between thumb and index finger
7. Open the transfusion line roller clamp
8. Purge the line of any air like during classic transfusion procedure
9.
Close the transfusion line roller clamp
10. Connect the transfusion line to the patient and
re-open the roller clamp
Flow regulation : through the drip chamber
operating mode
Essential precaution : perform the classic purge
2
4
Instructions for use - Version 2
1
5
7
9
3
6
8
10
10/15