Droper Field 1000 Instructions For Use Manual - Page 7
Browse online or download pdf Instructions For Use Manual for Medical Equipment Droper Field 1000. Droper Field 1000 15 pages. Mechanical pressure infuser
Also for Droper Field 1000: Operating Instructions Manual (16 pages)
End of administration
The end of the administration can be checked by :
observing the inverted pyramid drawn on the front surface of the mobile plate of the
equipment when the final peak will be visible.
observing, thanks to the cover transparency, the full crushing of the delivered flexible bag.
coupled with the absence of flow of infusion solution in the drip chamber if it is used.
if necessary, by observance of blood reflux in the tubing (with the device being located below
the level on which the patient is lying).
Once administration is completed, close the roller clamp of the infusion / transfusion line, open the
Droper Field using the same method as described in point Opening the cover, and remove the empty
bag from the device. Handle then the flexible bag and the infusion / transfusion lines wastes as when
with classical delivery by gravity or with pressure infusion cuffs.
Stopping administration
This can be stopped by :
closing the roller clamp or flow regulator by turning them to the OFF position.
opening the device (see point Opening the cover), the Droper Field being then at the same
level as the patient
means of any other suitable device.
Necessary accessories
The Droper Field does not need any dedicated accessories. It works with the flexible bags available in
the market up to 1000 ml and the existing infusion / transfusion lines.
:
7. Infusion
Two types of infusion lines are possible :
Classic infusion lines with drip-chamber, such as those used for gravity infusion or with
pressure infusion cuffs.
Or
Infusion lines with « Air Stop » filter (Intrafix SafeSet™, AirCatch™,...). These lines are
equipped with a hydrophilic membrane which prevents any air flow from the drip chamber
to the patient.
choice of line and operating mode
Instructions for use - Version 2
7/15