Beckman Coulter UniCel DxC 600 Daily Start Up - Page 5

Browse online or download pdf Daily Start Up for Medical Equipment Beckman Coulter UniCel DxC 600. Beckman Coulter UniCel DxC 600 15 pages. Synchron clinical system

Beckman Coulter UniCel DxC 600 Daily Start Up
Clearing Racks does not clear programming for Sample ID's; it only
clears the assignment to a rack/position. Your Closed Tube Tracking
Database tracks which samples have been pierced by Sample ID.
You may get a message informing you that some racks are In-Progress,
Removed, Requested for calibration, or Reserved for Cal. This feature
protects you from inadvertently clearing a rack that is needed for one of
these conditions.
UniCel DxC, Customer, Daily Start Up Job Aid
Rev. F (September 2013)

Clear Manually Assigned Racks

Select Samples from the menu bar
Select Clear F7
Select the Rack(s) field. Type the range of
rack numbers(s) to be cleared
Note: Enter 1-999 to clear all racks
Select Enter
from the keyboard
Select OK
Select OK
to confirm
Notes about Clearing Racks
Beckman Coulter Professional Development
Page 5 of 15