Ankom Daisy D200 Replacement - Page 4

Browse online or download pdf Replacement for Accessories Ankom Daisy D200. Ankom Daisy D200 4 pages. Motor replacement

11.
120v Motor: Reconnect the four motor wires to the Terminal Block
Assembly.
a. Twist the wire ends to prevent fraying as they are reinserted into
their designated slots on the terminals (as shown in the above
image for the 220v Motor).
b. Place the wire ends in their correct terminal blocks one at a time.
Using the small flat-blade screwdriver, screw the blue and white
wires into the #1 gray terminal block and the brown & black
wires into the #3 blue terminal block.
c. Gently pull each wire to make sure they are secure in the terminal.
12.
120v Motor: Fan installation.
Install the fan, which was sent with the new 120v motor, onto the
motor shaft by pushing until it securely snaps into place.
13.
Test the motor.
a. Turn on the instrument and test the motor.
b. If the motor is running properly, proceed to reinstall the back panel of the instrument.
14.
If the motor QC Test fails:
a. Check to ensure that the above steps have been performed
correctly and that all parts are securely installed.
b. Perform a second motor test. If this fails, contact ANKOM
Technology by email at [email protected] or by phone
315.986.8090
15.
Reinstall the back panel.
Replace the back panel of the Daisy instrument securing it with
the four screws removed in step #1.
With this completed you are now ready to return your Daisy
instrument to use.
ANKOM Technology  2052 O'Neil Road  Email: [email protected]  Phone: 315-986-8090  Fax: 315-986-8091
Motor Replacement
D200/D200I
Service Procedure
Revised: 08/06/2014
Terminal #1
Terminal #3
001
120v
Motor
Shaft
pg. 4 of 4