Druck DPI 841 User Manual - Page 9
Browse online or download pdf User Manual for Test Equipment Druck DPI 841. Druck DPI 841 14 pages. Frequency calibrator and frequency loop calibrator
Operation - Supply Hz or pulses
To supply Hz or pulses:
1. Connect the instrument (Figure 2) and, if necessary,
adjust the Set Up (Table 1).
2. Select a Hz or Pulses output task from Select Task
(Table 2/3) and, if necessary, adjust the Settings
(Table 5).
For Pulses, the display shows the time (hh:mm:ss) to
complete the count at the specified rate.
3. Supply the output values to the system (Table 7).
Pulses
Figure 2: Example configuration - To supply Hz or Pulses
Operation - Transmitter calibration
DPI 842 only. To calibrate a transmitter:
1. Connect the instrument (Figure 3/4) and, if necessary,
adjust the Set Up (Table 1).
2. Select the applicable calibration task from Select Task
(Table 2/3) and, if necessary, adjust the Settings
(Table 4/5).
3. Supply the output values to the system (Table 7).
K395 Issue 3
Pulses
Edit
Settings
Start
a) Hz
b) Pulses
Figure 3: Example configuration - Transmitter calibration
Settings
Figure 4: Example configuration - Transmitter calibration
Maximum: 30V
with external loop power
with internal loop power
Edit
Settings
Edit
Settings
7
[EN] English -