Druck InstruMetrics ADTS542F Safety And Installation Manual - Page 7
Browse online or download pdf Safety And Installation Manual for Test Equipment Druck InstruMetrics ADTS542F. Druck InstruMetrics ADTS542F 16 pages. Air data test set
Note: The battery does not necessarily need to be new or re-charged. The ADTS Touch will
function safely and correctly with a flat battery when the umbilical cable or dock charging
supply is connected.
4.
Refit the battery compartment cover plate and tighten the two thumbscrews (5). The cover
must be fitted correctly to maintain protection afforded.
2.2 Battery Care and Maintenance
The battery unit contains a Lithium-ion battery which requires no user maintenance. If shipping
an ADTS Touch and/or battery unit please contact your local service center beforehand for
shipping requirements. Battery must be removed from the ADTS Touch during transportation.
Prolonged exposure to temperature extremes may significantly reduce battery lifetime. For
maximum lifetime, avoid prolonged periods where the battery is exposed to temperatures outside
the range -30°C to +45°C.
The recommended storage temperature range is 5°C to 21°C (41°F to 98.8°F).
The battery is removable. The manufacturer makes the following safety recommendations. The
user should not:
•
Short circuit the battery.
•
Immerse the battery in any liquid.
•
Disassemble or deform the battery.
•
Expose to, or dispose of the battery in a fire.
•
Subject the battery to excessive physical shock or vibration outside the specified limits for
the ADTS.
•
Use a battery that appears to have suffered abuse.
•
Charge with an unspecified charger or specified charger that has been modified. This can
cause breakdown of the battery or swelling and rupturing.
•
Use a battery in an appliance or purpose for which it was not intended.
The battery unit is sealed for life, so leakage of electrolyte is not expected. If leakage of electrolyte
from the battery is observed, stop using the battery immediately and avoid contact with the
electrolyte. If skin or clothes come into contact with the electrolyte, wash immediately with soap
and water. If electrolyte comes into contact with the eyes, wash the eyes thoroughly with water
and consult a doctor immediately.
The ADTS5xxF, ADTS Touch and ADTS Touch-ER contain a Lithium-ion coin cell. This coin cell
is not user replaceable. Installation and replacement shall only be performed by trained
personnel.
Always dispose of batteries safely and in accordance with local instructions.
2.3 Battery Charging
The battery in the ADTS Touch can be charged using one of the following methods:
•
By placing the ADTS Touch onto a powered ADTS.
•
By connecting the ADTS Touch connector (4) (see Figure 3) to the ADTS using the umbilical
cable.
•
By using the AATOUCH-5 optional accessory desktop charger. The battery must be
removed from ADTS Touch when it is charged by the AATOUCH-5 desktop charger.
Copyright 2014 Baker Hughes Company.
English–ADTS Safety and Installation Guide | 5