Drucker Horizon 24 Service Manual - Page 9
Browse online or download pdf Service Manual for Laboratory Equipment Drucker Horizon 24. Drucker Horizon 24 12 pages. Flex centrifuge
Fig 9
•
Place the base assembly onto the guard bowl assembly with the guard bowl splice
positioned as shown (fig 9).
•
Feed the motor wire harness into the base passthrough so that the wire harness is
not pinched between the base and the guard bowl (fig 10).
Fig 11
•
Secure the base with six #8 screws torqued to 10 in-lbs. (fig 11).
•
Install the exhaust cover and secure with three #8 screws torqued to 10 in-lbs. (fig
12).
DRUCKER DIAGNOSTICS
HORIZON 24 FLEX SERVICE MANUAL
SPLICE
THIS SIDE
Fig 10
Fig 12
SM029
REV A
9