Drucker Horizon 24 Service Manual - Page 9

Browse online or download pdf Service Manual for Laboratory Equipment Drucker Horizon 24. Drucker Horizon 24 12 pages. Flex centrifuge

Fig 9
Place the base assembly onto the guard bowl assembly with the guard bowl splice
positioned as shown (fig 9).
Feed the motor wire harness into the base passthrough so that the wire harness is
not pinched between the base and the guard bowl (fig 10).
Fig 11
Secure the base with six #8 screws torqued to 10 in-lbs. (fig 11).
Install the exhaust cover and secure with three #8 screws torqued to 10 in-lbs. (fig
12).
DRUCKER DIAGNOSTICS
HORIZON 24 FLEX SERVICE MANUAL
SPLICE
THIS SIDE
Fig 10
Fig 12
SM029
REV A
9