Impedance Measurements

Depress the Ohms button to go into ground impedance
mode. You should see a screen similar to the one
shown on the right. The impedance mode is indicated
by "OH" on the bottom display line.
Scaling
The 3M
Ground Pro Ground Integrity Meter has
four different scales for measuring ground impedance. Choose the appropriate scale
for your measurements using the Up and Down arrow buttons.
O hm s
0
2
4
6
8
Auto-Zeroing Test Leads
Test leads offer added impedance and may alter
measurement results, especially at a lower scale. In
order to take impedance of test leads into account,
short the ends of the test leads together and depress the
Function button (F) until you see the screen shown on
the right. In a few seconds, this screen will disappear
and you will hear a short beep. The Ground Pro will
return to normal measurement mode. Now when the
test leads are shorted together, the display should show
zero Ohms, even though the test leads may have some
impedance. The correction factor will be saved in non-
volatile memory.