Troubleshooting:

NOTE:
The latch must be turned completely clockwise to its stop position in order
for the centrifuge to operate.
Problem:
1.
Solutions:
Problem:
2.
Solutions:
Problem:
3.
Solutions:
For servicing information or additional technical support, contact
Drucker Diagnostics at
The rotor does not spin freely.
– Make sure nothing has fallen into the rotor chamber.
– If there is nothing obstructing the rotor, the rotor may be damaged
Contact Drucker Diagnostics for further assistance.
Excessive noise when the centrifuge is running.
– Check to see that the load is balanced.
– Make sure that nothing has fallen into the rotor chamber.
– Make sure that the nut in the center of the rotor is tight.
– Have a technician test the motor and replace it if necessary.
The centrifuge does not run.
– Check the electrical outlet.
– Make sure the lid latch is turned completely clockwise to its stop
position. When the lid is closed properly, the latch light on the control
panel will illuminate.
– Check the fuse in the end of the power cord. If the filament is broken,
the fuse has blown and must be replaced with a fuse of the same
type and rating. Contact Drucker Diagnostics for further assistance.
814-342-6205 or 814-692-7661