Drucker Diagnostics 614VET Operator's Manual - Page 13
Browse online or download pdf Operator's Manual for Laboratory Equipment Drucker Diagnostics 614VET. Drucker Diagnostics 614VET 16 pages.
T
ROUBLESHOOTING
Note: The latch must be turned completely clockwise to its stop position
in order for the centrifuge to operate.
The rotor does not spin
freely.
Excessive noise when
the machine is
running.
The centrifuge does
not run.
The latch light does
not come on when the
lid is closed.
Page | 13 Drucker Diagnostics +1-814-692-7761 - [email protected]
o
Make sure nothing has fallen into the
rotor chamber.
o
If there is nothing obstructing the rotor,
the rotor may be damaged. contact
Drucker Diagnostics for further
assistance.
o
Check to see that the load is balanced.
o
Make sure that nothing has fallen into
the rotor chamber.
o
Make sure that the nut in the center of
the rotor is tight.
o
Have a technician test the motor and
replace it if necessary.
o
Check the electrical outlet.
o
Make sure the lid latch is turned
completely clockwise to its stop
position. When the lid is closed
properly, the latch light on the control
panel will illuminate.
o
Check the circuit breaker switch at the
bottom left of the machine. If the switch
is white, the breaker has tripped.
Contact Drucker Diagnostics for further
assistance.
o
The printed circuit board may be
damaged. Have a technician test and
replace the circuit board if necessary.
o
Make sure that the unit has power.
o
Make sure the lid latch is turned
completely clockwise to its stop
position. The latch makes contact with a
switch underneath the front top of the
cabinet. If this switch is not activated,
the light will not turn on and the
machine will not run.