Drucker Diagnostics DASH Flex Series Operator's Manual - Page 12
Browse online or download pdf Operator's Manual for Laboratory Equipment Drucker Diagnostics DASH Flex Series. Drucker Diagnostics DASH Flex Series 18 pages.
O
PERATION
o Place the tubes into the tube holders. Be sure to follow the rules for balanced loads as listed in the next section.
o Close the lid and turn the lid knob clockwise to its complete stop position.
o The digital screen shows the currently selected cycle. To select another cycle, press the UP or DOWN button in
succession until the desired cycle is selected.
o Pushing the START button on the control panel will start the spin cycle.
o When the cycle is completed, the rotor will slow to a complete stop and the lid light will flash.
o The unlocking mechanism will engage for 60 seconds allowing entry into the rotor chamber. To unlock after more
than 60 seconds have elapsed, press the UNLOCK button. The lid will unlock for another 15 seconds.
o Turn the lid knob counterclockwise and open the lid. The lid light will turn off.
o You may now safely remove the samples.
B
L
ALANCING
OADS
Your centrifuge must contain a balanced load to work properly. Spinning balanced loads will extend the life of the
centrifuge and produce better results. Use the following rules when loading the rotor.
is to be spun, fill a tube with water to match the weight of the unpaired sample and place it across from this sample.
Opposing tube holders must be equally loaded or empty or loaded with equally weighted samples. When loading
only 3 tubes, they must be of equal weight.
12-T
C
UBE
ENTRIFUGES
Drucker Diagnostics – Customer Service: +1-814-692-7661 – [email protected] | Page 8
f an odd number of samples
I