Anritsu Company ShockLine MS46122B Series Quick Start Manual - Page 5
Browse online or download pdf Quick Start Manual for Measuring Instruments Anritsu Company ShockLine MS46122B Series. Anritsu Company ShockLine MS46122B Series 20 pages. Vector network analyzers verification kits and performance verification software
4.
PC Controller Cable Connections to VNA
MS46524B Verification Component Connection
The basic connections between the PC Controller, the MS46524B VNA, and the verification components are
shown below. See
Figure 2
Port-4 are verified as a set.
8
7
1. ShockLine Series VNA
2. PC Controller:
• With installed NI VISA Library
• With installed 3-2300-610 PVS application
3. Ethernet connectivity
4. Verification Kit USB Device with characterization data
Figure 2. MS46524B VNA Connections
MS46522B Verification Component Connection
The basic connections between the PC Controller, the MS46522B VNA, and the verification components are
shown below. See
Figure 3
1. ShockLine Series VNA
2. PC Controller:
• With installed NI VISA Library
• With installed 3-2300-610 PVS application
3. Ethernet connectivity
4. Verification Kit USB Device with characterization data
Figure 3. MS46522B VNA Connections
3-2300-610 Verification Kits/PVS QSG
below. Note: For the MS46524B: Port-1/Port-2 are verified as a set and Port-3 and
1
5
below.
1
5
8
7
PN: 10410-00766 Rev. B
4. PC Controller Cable Connections to VNA
3
6
5. Port 1 Adapter (if needed)
6. Test Port Through Line M-F Cable
7. Adapter attached to Through Line and Port 2
8. Insert calibration or verification components here.
Note that the setup for any given verification kit varies
slightly with additional adapters.
3
6
5. Port 1 Adapter (if needed)
6. Test Port Through Line M-F Cable
7. Adapter attached to Through Line and Port 2
8. Insert calibration or verification components here.
Note that the setup for any given verification kit varies
slightly with additional adapters.
2
4
2
4
QSG-5