Anritsu Company 34NFNF50 Calibration Manual - Page 13

Browse online or download pdf Calibration Manual for Adapter Anritsu Company 34NFNF50. Anritsu Company 34NFNF50 16 pages. Precision coaxial adapters

Performance Verification
9. Under the Set Scale Key:
a. Set the Resolution to 500 mU/div.
b. Set the Reference Value equal to 1.1
10. Install the Adapter Under Test to VNA Port 1.
11. Install the appropriate termination from the 3653 cal kit to the
open end of the Adapter Under Test.
12. Verify that the SWR of the Adapter Under Test is within
specification (refer to
Hint: Use a marker to find the maximum SWR value.
Model 34 Series CG
Table 2-1, "Model 34 Series
PN: 10100-00057 Rev. B
2-3 Test Procedure
Specifications).