2B Technologies 211 Manual - Page 3

Browse online or download pdf Manual for Measuring Instruments 2B Technologies 211. 2B Technologies 211 7 pages. Portable n2o source
Also for 2B Technologies 211: Quick Start Manual (2 pages)

2B Technologies 211 Manual
Figure 1. Diagram of the Portable N
To install the Portable N
1) If there is no regulator mount on the instrument, first install the mount.
Model 211: If you have a newer instrument that has a regulator mount on the back
panel, proceed to Step 2. Otherwise, a new back panel will need to be requested
from 2B Tech and installed. Remove the top and bottom cover first and disconnect
all connections to the back panel. Taking a photo of the connections to the back
panel is a good idea to help ensure that the connections are remade correctly after
installation of a new back panel. Transfer all the components on the old back
panel to the new one. Mount the regulator mount to the back panel using the two
#10 pan head screws and split locking washers supplied with the Portable N
Source. Install the back panel on the instrument and install the top and bottom
panels.
O Source follow the following steps:
2
2
O outlet (1/8" barb)
N
2
Pressure regulator
On/off valve (black knob)
Inlet pressure gauge
Two #10-32x1/2" Phillips head
screws with split washers
Regulator mount
One #10-32x1" socket cap screw with
flat washer
Removable cartridge cup to hold
disposable N
removable insert enables the use of
either 8-g or 16-gram cartridges.
O Source and mounting hardware.
2
Portable N
O cartridges. A
2
O
2
O Source Manual Rev. B-2
2