Fault
The scaling tip vibrates
but there is no spray when
stepping on the foot switch.
The vibration of the tip
becomes weak.
There is water seeping from
the coupling between the
handpiece and cable.
There is water flow out when
8
turn off the power.
The handpiece generates
heat.
The amount of spouting
water is too little.
The vibrating intensity
control knob is seized up.
The u-file doesn't vibrate.
There is noise coming from
the endochuck.
LED light don't work
Possible cause
The water control knob is not on.
The tip hasn't been screwed on to the handpiece tightly.
The tip is loose because of vibration.
The coupling between the handpiece and the cable isn't dry.
The tip is damaged [note 2].
The waterproof "O"ring is damaged.
There is impurity in the solenoid valve.
The amount of spouting water is too little.
The potentiometer is broken.
The water control knob is a low grade.
The water pressure is not enough.
The water pipe is jammed.
The potentiometer is damaged.
The screw is loose.
Endochuck is damaged.
The screw is loose.
Poor contact
Something wrong with LED light
Solutions
Turn on the water contrl knob [note 1].
Screw the tip on the handpiece tightly (See Picture 4).
Screw on the tip tightly (See Picture 4).
Dry it by the hot air.
Change a new one.
Change a new waterproof "O"ring.
Contact with the local distributor or manufacturer.
Turn the water control switch to a higher grade [note 1].
Change a new one.
Turn the knob to a high grade [note 1].
Enhance the water pressure.
Clean water pipe with multi-function syringe [note2].
Contact with the local distributor or our company.
Tighten it.
Change a new one.
Tighten it.
Contact tightly
Change a new one