8. Precaution

8.1 Please recharge the battery at least 4 hours before first time usage.
8.2 During operation, the light should be aimed straightly at the resin to
ensure the effect of solidification.
8.3 Avoid aiming the blue light at eyes directly.
8.4 Only the original pedestal charger, adapter and Lithium battery could
be used, because other brand pedestal charger, adapter and Lithium
battery are likely to damage the circuit.
8.5 It is forbidden to touch the charging connector with metal or other
conductor, to avoiding damage the circuit of charge or the battery.
8.6 Charging the battery in the condition of cool and ventilated. Please
make sure of pressing out the buckle between the main unit and the
pedestal, otherwise the battery charging might be failed because of the
poor contact.
8.7 Do not disassemble the Lithium battery, it will lead to the circuit
short or the electrolyte leakage.
8.8 Do not squeeze, shake and short the battery, do not store the battery
with metal material.
① WARNING: If the curing light works for 40s continously, the
temperature of the top of optical fiber may reach 56 ℃ ."
② WARNING: Do not modify this equipment without authorization
of the manufacturer.

9. Contraindication

The heart disease patients, pregnant women and children should be
cautious to use this equipment.

10. Maintenance

10.1 The optical fiber should be sterilized for 4 minutes with 134 ℃ and
6