Auber Instruments AT210-A Notes On Use - Page 2

Browse online or download pdf Notes On Use for Thermometer Auber Instruments AT210-A. Auber Instruments AT210-A 2 pages.

Auber Instruments, Inc.
www.auberins.com
For example, two of the thermometer (AT210-A) units have their ID number set as "001" and "002" respectively.
The remote unit (AT210-B) is currently receiving temperature reading from thermometer unit "002". So, the
thermometer unit "001" is currently NOT paired with the monitor (AT210-B). When this thermometer unit "001"
is switch off and then switched on again, the temperature reading jump to around 2223°F with the sensor
plugged in (Note: press the SHIFT ">" key the mute the buzzer.), or show -20°F if the sensor is unplugged.
However, this error is temporary and it can be fixed by following the steps below.
To fix this issue:
1) Switch OFF the monitor unit (AT210-B) by unplug its power cord.
2) Then, cycle the power to the thermometer (AT210-A) unit whose reading is erroneous. The temperature
reading should go back to normal.
3) Switch ON the monitor unit again by plugging in the power cord. Now, the system should work correctly as
before.
Special Note
The AT210 system is designed to use one monitor unit (AT210-B) to read from one or more thermometers
(AT210-A). Other combinations are NOT recommended NOR supported. Using two monitors (AT210-B) and two
thermometers (AT210-A), i.e., two sets of AT210-COMBO, can result in communication failure and all units won't
be able to function properly.
(However, in only one special case, where there are two monitors (AT210-B) and only one thermometer (AT210-
A), the system can work fine most of the time. Both monitors can receive temperature reading from the
thermometer, but occasionally they will lose connection and show "----". Usually, the readings will repopulate
shortly. )
Please
contact us
for any question during the troubleshooting process.
(End)
Auber Instruments
5755 North Point Parkway, Suite 99
Alpharetta, GA 30022, USA
www.auberins.com
E-mail: [email protected] Tel: 770-569-8420
Copyright 2007-2022, Auber Instruments. All Rights Reserved.
No part of this manual shall be copied, reproduced, or transmitted in any way without the prior, written consent of Auber Instruments. Auber
Instruments retains the exclusive rights to all information included in this document.