couplant or cable attenuation. C3 is also ideal for thin materials as well as materials such as
steel, aluminum and ceramics where multiple echoes exists.
Class 5 or the
A2-DFR CLF4 transducer setup found in the transducer menu
of the
5=
ECHO 7/Precision meets Boeing's specification to inspect a range from .007"- 1.0" with
one transducer and one setup. (requires appropriate transducer)
.
Description of Classes of Measurements:
DEFP2- DLK1025 is a default class 2 set up for P for Plastics with a 10 Mhz, .25 inch probe.
DEFS is for Steel and the classes of measurement are either 1, 2 or 3. Class one uses a contact
transducer to measure the time from initial pulse or main bang to first backwall echo. Class 2 is
uses a delay or immersion probe and measures the time from the interface echo to first back wall
echo. Class 3 also uses a delay line or immersion probe and measures the time between multiple
echoes after an interface echo.
Class 1 with cursor on first backwall echo
Ver: 2.0