Additel 918 User Manual

Browse online or download pdf User Manual for Water Pump Additel 918. Additel 918 2 pages. Pneumatic pressure test pump

Pneumatic Pressure Test Pump User Manual
Configuration & Air routine
1
2
3
5
7
6
8
Under Vacuu
Troubleshooting
Problem
Hand pump will not generate pressure
It is difficult to increase pressure
Hard to use the fine adjust
It is difficult to turn the quick connector
917/918
's
[Version number:1407V02]
1 - Hand pump
2 - Transport handle
3 - Quick connector
4 - Vent valve
(clockwise to close/ counterclockwise to open)
5 - P/V selector lock
6 - Pressure/vacuum (P/V) selector
Positive pressure: pull it out; vacuum: push it in
!
DO NOT switch under pressure or vacuum to
avoid damage)
4
7 - Isolation valve (Bypass valve)
(Isolated the calibration volume from the hand pump,
piston, and check valve)
8 - Fine adjust handle
(clockwise to increase pressure)
Cause
The isolation valve is not open
.
A. Vent valve is not closed.
B. The O-ring seal is loose or damaged.
C. P/V selector is in the wrong position
A. The isolation valve is not closed
.
B. The gauges are not tightened.
C. The O-ring seal is damaged.
D. The thread surface is not smooth.
E. The connector type mismatched to the gauge pressure port.
F. Dust or contaminants are blocking the pressure flow
A. Too much force was previously applied.
B. The threads have no lubrication
.
Warnings and cautions
Low pressure gauges can very easily be over pressured if not
careful.
Please take caution when applying pressures.
Do not exceed the safety limit of 1500 psi (100 bar) for 917 or
2000 psi (140 bar) for 918.
Do not switch between pressure and vacuum mode under
pressure. Open the vent valve and isolation valve before
switching between positive to vacuum pressures.
Over tightening connectors may cause damage. Light hand
tightening is sufficient.
Store the pump in a dry non-corrosive environment.
Additel is not responsible for problems or damage resulting from
improper use or operation of the equipment.
All moisture and contaminants should be cleaned out of the liquid
isolator before creating a vacuum. This is accomplished by
venting the pump when at high pressure.
For the least risk in contamination of the reference gauge, the
DUT should be installed to the pressure port which is closest to
the vent valve.
Specification
Pressure range:
Temperature:
Humidity:
< 85%RH
Adjustment resolution
Maximum pressure:
Pressure media:
Size:
Height: 5.5'' ( 140 mm );
Base: 12.4'' ( 315 mm ) x 7.8'' ( 198 mm )
Weight:
5.7 lb ( 2.6 kg )
Remark:If local atmosphere pressure is 1 bar, the vacuum can reach to -0.95 bar;
If local atmosphere pressure is P, the vacuum can reach to - 0.95P.
Open the isolation valve
Close the vent valve.
Replace O-ring seal
.
Pressure: pull the P/V selector out; Vacuum: push the P/V selector in.
Close the isolation valve
Tighten the reference gauge or the gauge under test
Replace the O-ring seal.
Use Teflon tape on the thread and turn it in tight
Use the correct adapter
Pressurize the pump several times and then suddenly vent which will push any
contaminants out.
Secure the quick connect valves with less force.
Apply lubrication to the threads
917: 95% vacuum to 1000 psi ( 70 bar )
91 : 95% vacuum to 1500 psi ( 100 bar )
8
0~50
)℃
:
10 Pa ( 0.1 mbar )
917: 1500
psi (
100
bar )
918
: 2000 psi ( 140 bar )
Air
Air routine
7
6
1
Solution
.
.
.
.
.
.
3
4
8