Audiomatica CLIO FW-02 Installation Manual - Page 11

Browse online or download pdf Installation Manual for Recording Equipment Audiomatica CLIO FW-02. Audiomatica CLIO FW-02 15 pages.

Audiomatica CLIO FW-02 Installation Manual

3.7 SYSTEM CALIBRATION

This section describes how to perform the system calibration.
Be sure that, any time you perform a calibration, the system has warmed up for, at
least 15-20 minutes.
Select Calibration from the Main menu ;
Leave the CLIO Box front plugs unconnected.
Answer OK to the initial prompt; this will run an automatic procedure that will last
several minutes. The calibration procedure is completely automatic and several
progress indicators will accompany all the executed measurements. At the end of it
your CLIO system should be calibrated and ready to carry out measurements.
Software
won't
allow
calibration
if
internal
temperature
lies
outside
a
predetermined range (between 30 to 45 degrees).
So it is normal to receive a message like:
It is also normal to receive the following message:
This is normally due to the calibration parameters that do not match the internally
stored parameters that have been measured at Audiomatica laboratory and that
ensure the highest precision of the instrument. In such a case either wait to repeat
calibration if instrument is too cold for it to warm, or wait for the device to cool if
too hot.
At the end of the calibration process it is always mandatory to verify the calibration
itself; this is done by two simple measurements as described in the following
3 CLIO INSTALLATION
11