Aplisens SGE-25S.SMART User Manual - Page 17
Browse online or download pdf User Manual for Accessories Aplisens SGE-25S.SMART. Aplisens SGE-25S.SMART 20 pages. Smart level probes
Also for Aplisens SGE-25S.SMART: User Manual (20 pages)
10. MAINTENANCE
10.1. Periodic inspections
Periodic inspections should be carried out in accordance with the standards applicable to the user.
Check the processing characteristics by following the steps appropriate to the calibration and
configuration.
10.1.1. External overview
During the inspection, check the condition of the separating membranes (tarnish, corrosion) and the
electrical connection (check the condition of the cable), as well as the stability of the handle (if used).
Check for signs of mechanical damage in the form of impact marks or dents.
10.1.2. "Zero" check
Every 2 years or in accordance with the user's standards, check the probe's "zero" by pulling the probe
above the liquid mirror and reading the output signal.
10.2. Non periodic inspections
If the probe has been subjected to mechanical damage, pressure overload, hydraulic impulses,
electrical surges, deposits, crystallization of the medium, etching of the diaphragm, or if the probe is
found to be working incorrectly, the device should be inspected. Check the condition of the diaphragm,
clean it, check the electrical functionality and processing characteristics.
If the signal is missing in the transmission line or its value is incorrect, check the power
supply line, the state of connections on terminal strips, connections, etc. Check the correct
value of the supply voltage and the resistance of the load.
10.3. Cleaning/Washing
To remove impurities from the external surfaces of the probe wipe it with a cloth dampened in water.
10.4. Diaphragm cleaning
The only possible method of cleaning the probe diaphragms is to dissolve the sludge produced.
Do not remove deposits and impurities from the probe diaphragms, which are formed
during operation, mechanically using tools, since the diaphragms and the probes can be
damaged.
10.5. Spare parts
Parts of the probe that may be worn or damaged and must be replaced: cable and gland seals. The
cable and seals can only be replaced by the manufacturer.
10.6. Repair
Faulty or non-operational probe shall be provided to the manufacturer or an authorized representative.
10.7. Return
In the following cases the, probe should be returned directly to the manufacturer:
-
need for repair;
need for factory calibration or „zero" calibration;
-
-
replacement of improperly selected/shipped probe.
Revision 01.A.001/2021.09
EN.IO.SGE.25.SMART
17